Applies to: Dental machines

Exchanging the main fuse

 

The internal power supply of the machine has a main fuse that is accessible from the outside and can be exchanged if necessary.

  • As a replacement fuse, only use a fuse of the following type: T6,3A L250V

A new main fuse is available as spare part from customer service.

  1. DANGER!Switch off the machine at the main power switch.

  2. Remove the power cord from the connection panel.
  3. Remove the cover (marked orange) of the fuse.
  4.  

  5. Remove the defective fuse and replace it with a new fuse.
  6. If you do not have a replacement fuse ready, take the replacement fuse from the right side of the fuse cover and put it into the left side of the fuse cover.
  7. Remount the fuse cover.