Exchanging the working chamber door
Watch the video
YouTube video – When viewing this video, personal data is sent to YouTube, LLC, USA. Privacy statement
-
Have ready:
-
Open the working chamber door so that the lower working chamber door handles (marked green) are aligned parallel to the machine floor.
-
Hold the working chamber door in place. Use the 1.5 Nm torque screwdriver to loosen the 4 screws (marked orange) that secure the working chamber door to the working chamber door bracket.
-
Remove the working chamber door.
-
Position the spare working chamber door so that the 4 screw holes on the rear of the working chamber door (marked orange) align with the 4 screw holes in the working chamber door bracket (marked green).
-
Using the 1.5 Nm torque screwdriver, screw down the working chamber door with the 4 self-tapping screws (marked orange).
-
Test if the working chamber door opens and closes properly. If not, loosen the 4 screws and correct the position of the working chamber door.