Applies to: E5

Exchanging the working chamber door

Watch the video

YouTube video – When viewing this video, personal data is sent to YouTube, LLC, USA. Privacy statement

  1. Have ready:

    • 1 x Torque screwdriver 1.5 Nm with TX20 bit

       

    • 1 x Spare working chamber door

  2. Open the working chamber door so that the lower working chamber door handles (marked green) are aligned parallel to the machine floor.

     

  3. Hold the working chamber door in place. Use the 1.5 Nm torque screwdriver to loosen the 4 screws (marked orange) that secure the working chamber door to the working chamber door bracket.

  4. Remove the working chamber door.

  5.  

     

     

  6. Position the spare working chamber door so that the 4 screw holes on the rear of the working chamber door (marked orange) align with the 4 screw holes in the working chamber door bracket (marked green).

  7.  

     

     

  8. Using the 1.5 Nm torque screwdriver, screw down the working chamber door with the 4 self-tapping screws (marked orange).

  9. Test if the working chamber door opens and closes properly. If not, loosen the 4 screws and correct the position of the working chamber door.

  10.